Ultrahigh field 7.0 T head-only MRI scanner

eagle-i ID

http://harvard.eagle-i.net/i/00000130-9196-128d-1971-4f2080000000

Resource Type

  1. MRI Scanner

Properties

  1. Resource Description
    "(This instrument has an) 80mT/m head gradient set and an Avano whole body gradient set. The 7.0 T 90 cm (magnet ID) whole-body magnet was built by Magnex Scientific (Oxford, UK), and the conventional MRI console, gradient drivers and patient table were provided by Siemens. Integration of these components and the design and construction of RF coils were performed by MGH and Siemens personnel. The 7T whole body magnet is augmented with a 80mT/m, 800 T/m/s slew rate head gradient set (Siemens) for echoplanar imaging and 2nd and 3rd order resistive shim coils under computer control. With its high performance gradient set, the system can provide better than 100-µm resolution. The system is shielded by a 460-ton steel shield."
  2. Contact
    Wald, Larry, Ph.D.
  3. Manufacturer
    0000013d-bc83-21a4-ecc6-81a580000000
  4. Related Technique
    Magnetic resonance imaging
  5. Location
    Martinos Center for Biomedical Imaging Core Facility (MGH)
 